3.2 Systems supplied

761 SD Compact IC / Instructions for Use 8.

761.1043

61

Links tab

Always connect the 766 IC Sample Processor to COM2.

Linked method

Method

SD_phosphate.mtw

is linked to system

auto.smt

.

Most important settings:

Always ensure that the path for

Chromatogram directory

on the

Processing tab of the

METHOD SETUP

window exists. Other-

wise, it must be adapted (see Section 3.1.3). The default path is

C:\Metrohm\SD Analyzer..

Fields

IC Cap

and

Excel

should be activated on the Export tab of

the

METHOD SETUP

window.

"4

min

"

should be entered for

Duration

on the General tab of the

METHOD SETUP

window.

3.2.4 System

"shutdown.smt"

System

shutdown.smt

is used after the actual measurements (at the end

of the working day) in order to shut down the instrument. Water is in-

jected and the runtime of the program is 7 minutes.

Always inject water as the last sample and thus rinse the tubing con-

nections with water in order to avoid damage to the instrument.