2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

20

2.6.7 Connection

suppressor

waste

Route the "suppressor outlet capillary for eluent ("Waste")" 44 at the rear

out of the 761 SD Compact IC into a waste container.

2.7

Installing the regenerant path

The regenerant (100 mmol/L H

2

SO

4

) is aspirated with the peristaltic

pump from the regenerant MPak and forced into the suppressor via the

filter unit PEEK. From the suppressor, it is then routed into a waste con-

tainer (see Figure 9).

2.7.1

Fitting the pump tubing for regenerant

One channel of the two-channel peristaltic pump is used to pump the

regenerant. Fit a pump tubing 61 6.1826.110 as follows:

1 Disengage the tubing cartridges

Detach and disengage the two tubing cartridges 57 above

the pump drive 49 by pressing in the snap-action lever 59 on

the retaining bracket.

Press the pressure lever 58 down fully.

2 Fit the couplings 60

Fit one coupling 60 6.2744.110 onto each end of the pump

tubing 61 6.1826.110. Moistening the tip of the coupling

lightly will make it easier to slip it on.

3 Fit the pump tubing

Insert the fitted pump tubing 61 as shown in Figure 8 into the

tubing cartridge. The left-hand stopper 62 should latch into

the corresponding fixture on the left-hand side of the tubing

cartridge.

4 Reengage the tubing cartridge

Reengage the tubing cartridge with pump tubing 61 in the re-

taining bracket. Engage it at the left first and then press the

right-hand side down until the snap-action lever 59 engages.

Ensure that the pump tubing is not kinked when doing this.

Leave the other tubing cartridge outside. It is fitted with the

pump tubing for the sample path (see Section 2.8.1).