3.2 Systems supplied

761 SD Compact IC / Instructions for Use 8

.761.1043

55

Configuration tab
Same settings as in the example with system

manual.smt

in Section

3.1.6.

Links tab
Same settings as in the example with system

manual.smt

in Section

3.1.6. The 761 SD Compact IC should always be attached to COM1.

Linked method

Method

SD_startup.mtw

is linked to system

startup.smt

.

Most important settings:

You should always ensure that the path for

Chromatogram directory

exists on the Processing tab of the

METHOD SETUP

window.

Otherwise, it must be adapted (see Section 3.1.3). The default

path is C:\Metrohm\SD Analyzer..

Field

IC Cap

should be activated on the Export tab of the

METHOD SETUP

window.

"30

min

"

should be entered for

Duration

on the General tab of the

METHOD SETUP

window.

3.2.2 System

"manual.smt"

System

manual.smt

is used to manually measure samples (see also

Section 5.1).
System

manual.smt

was used in Section 3.1 as an example for the

screenshots.

System window

See Section 3.1.2.

System parameters

See Section 3.1.6.

Linked method

Method

SD_phosphate.mtw

is linked to system

manual.smt

.

Most important settings:

You should ensure that the path for

Chromatogram directory

on the

Processing tab of the

METHOD SETUP

window exists. Other-

wise, it must be adapted (see Section 3.1.3). The default path is

C:\Metrohm\SD Analyzer..

Fields

IC Cap

and

Excel

should be activated on the Export tab of

the

METHOD SETUP

window.

"4

min

"

should be entered for

Duration

on the General tab of the

METHOD SETUP

window.