Connection sample vessel, Pump tubing, Injection valve – Metrohm 761 SD Compact IC User Manual

Page 32: Waste, Figure 10, Pump tubing for sample path, Peek coupling 6.2744.030, Pump tubing 6.1826.110, Peek coupling 6.2744.160, Fit the coupling 63

Advertising
background image

2 Installation

761 SD Compact IC / Instructions for Use 8.

761.1043

24

Figure 10: Pump tubing for sample path

57 Tubing cartridge

58 Pressure lever

59 Snap-action lever

62 Stopper orange-yellow

63 PEEK coupling 6.2744.030

64 Pump tubing 6.1826.110 for sample

65 PEEK coupling 6.2744.160

with tubing security device

2.8.2

Connection sample vessel

pump tubing

injection valve

waste

Proceed as follows in order to make the connections for the sample

path:

1 Connection Sample vessel - pump tubing

The sample is aspirated with the aspirating tubing 3:

The aspirating tubing 3 is routed through the "feedthrough for

aspirating tubing" 2 out of the front door into the sample ves-

sel. It is fixed in position on the feedthrough 2 with a rotary

nipple (see Section 2.9.2 if using the 766 IC Sample Proces-

sor).

Detach the other end of the aspirating tubing 3 with pressure

screw 50 from the injection valve (it is fitted there only for de-

livery purposes) and screw it firmly onto the PEEK coupling

63 on the left-hand end of the pump tubing (flow inlet) (see

Figure 11).

65

64

62

59

63

57

58

62