2.7 Installing the regenerant path

761 SD Compact IC / Instructions for Use 8.

761.1043

21

Figure 8:

Pump tubing for regenerant path

57 Tubing cartridge

58 Pressure lever

59 Snap-action lever

60 Coupling 6.2744.110

61 Pump tubing 6.1826.110 for H

2

SO

4

62 Stopper orange-yellow

2.7.2 Connection

regenerant-MPak

pump tubing

suppressor

waste

The connections to the pump tubing, suppressor and waste must be

connected first. Then, and not before, connect the end at the MPak

outlet. If the end at the MPak outlet is connected first, the regenerant

(H

2

SO

4

) will begin to escape.

Proceed as follows:

1 Connection Pump tubing - Suppressor

Screw the filter unit PEEK 56 6.2821.120 onto the coupling 60

6.2744.110 at the right-hand end (flow outlet) of the pump

tubing 61 fitted in the peristaltic pump. Note the flow direction

arrow on the filter unit.

Firmly screw the "suppressor inlet capillary for regenerant

("H2SO4")" 45 onto the filter unit 56 with a pressure screw 50

6.2744.010 (see also Section 2.3.3).

2 Connection Suppressor - Waste

Route the "suppressor outlet capillary for regenerant

("Waste")" 46 at the rear out of the 761 SD Compact IC into a

waste container.

58

60

62

57

60

61

59

62