6.2 Maintenance and servicing

761 SD Compact IC / Instructions for Use 8.

761.1043

91

6.2 Maintenance

and

servicing

6.2.1 General

information

Care
The 761 SD Compact IC requires proper care and attention. Excessive

contamination of the instrument could possibly lead to malfunctions

and a shorter service life of the inherently rugged mechanical and

electronic parts.
For protection against escaping fluids, the two drain tubes for the inner

compartment and for the bottle rack must be fitted (see Section 2.5).
Spilled chemicals and solvents should be wiped up immediately. It is

especially important to protect the plug connections at the rear of the

instrument (in particular, the mains plug) against contamination.

Even though design measures virtually eliminate such a situation,

should corrosive media penetrate the interior of the instrument, the

mains plug of the 761 SD Compact IC must be immediately discon-

nected to prevent extensive damage to the instrument electronics. In-

form Metrohm Service if your instrument(s) has or have been dam-

aged in such a way.

The instrument may not be opened by untrained personnel. Please

comply with the safety notes in Section 1.4.

Maintenance by Metrohm service
Maintenance of the 761 SD Compact IC is best done as part of an an-

nual service performed by specialists from the Metrohm company. If

work is frequently performed with caustic and corrosive chemicals, it

may be necessary to shorten the interval between servicing.
The Metrohm servicing department is always willing to offer expert ad-

vice on the maintenance and servicing of all Metrohm instruments.

6.2.2 Passivation

Passivation of the complete IC system (without column) by rinsing with

20...50 mL 0.2 mol/L HNO

3

is required only if unusual changes to the

measuring properties of the cell are observed. In this case, remove the

separating column 36 from the 761 SD Compact IC and connect the

two capillaries 35 and 37 (see Figure 3) directly together using a cou-

pling 54 (6.2744.040).

6.2.3 Shutdown

If the 761 SD Compact IC is shut down for a considerable length of

time, the entire IC system (without column and suppressor) must be

Advertising