2.13 Deaerating the pump and rinsing the pulsation dampener

761 SD Compact IC / Instructions for Use 8.

761.1043

35

3 Open the control window

Double-click on the 761 image in the System window. You will

see the window for manual operation of the 761 SD Compact

IC (see below).

4 Deaerate the pump

Ensure that the "tubing connection to the eluent MPak" 23 is

connected to the eluent MPak.

Click on the

<On>

button for

IC pump

to switch on the high-

pressure pump.

Aspirate air with the syringe inserted in connection 6 until elu-

ent flows into the syringe.

Click on the

<Off>

button for

IC pump

, to switch the high-

pressure pump back off.

Close the rotary knob on the purge valve 26 by turning it

clockwise.

Remove the syringe from the connection 6.

2.13.2 Rinsing the pulsation dampener

The pulsation dampener must be rinsed before using it for the first time

so as to remove the isopropanol which is located in the flow area of the

eluent on delivery.

The isopropanol should not be rinsed through the suppressor and the

detector. Consequently, the column connection capillary

33

(which is

currently connected to the suppressor via the PEEK coupling) must be

detached from the PEEK coupling and routed into a beaker.