5.1 Operation with manual sample change

761 SD Compact IC / Instructions for Use 8.

761.1043

77

5 Operation

This section describes day-to-day operation of the 761 SD Compact

IC. A distinction is made between operation with manual sample

change (Section 5.1, Instrument version 2.761.0420) and operation

with automatic sample change with the 766 IC Sample Processor

(Section 5.2, Instrument version 2.761.0520).

5.1

Operation with manual sample change

The instrument must have been connected and the related software

must have been installed as described in Section 2 before the instru-

ment can be operated.

Switching the equipment on:

1 Switch on the PC

Switch on the PC.

2 Switch on the 761 SD Compact IC

Switch the 761 SD Compact IC on with the mains switch 14

on the rear panel of the instrument. The pilot lamp 7 lights af-

ter the instrument has been switched on.

5.1.1

Control with «IC Net» (manual operation)

Launch the «IC Net» software and log in (see Section 2.12.3).

5.1.1.1

Conditioning (manual operation / «IC Net»)

The instrument should be conditioned each time it is restarted:

1 Open and link system startup.smt

Open (see Section 3.1.2) system

startup.smt

(see Section

3.2.1) and link it (see Section 3.1.4).