3.3 Fitting the accessories

838 Advanced Sample Processor, Installation

17

3.3.6 Fitting the ultrafiltration cell

This section describes initial fitting of the ultrafiltration cell including fil-

tration membrane. Please note Chapter 6.1 when assessing the need to

exchange the membrane and when selecting membranes. The proce-

dure for exchanging a filtration membrane is also described in full in

Chapter 4.2.2.

Figure 17 Fitting the ultrafiltration cell


28

Ultra filtration cell (6.2729.110)

33 Outlet for filtrate

29

Upper section of the filtration

cell

34

Filtration membrane

(6.2714.020)

30 Sealing ring (E.301.0111)

35

Inlet for sample solution

31

Lower section of the filtration

cell

36 Outlet for sample solution

32

Screws (V.022.6030) including

washer (4.754.4090)

1 Preparing the ultrafiltration cell

• Remove the ultrafiltration cell 28 from its packaging and remove the

three screwed-on blind stoppers.

• Using the 6.2621.070 Allen key (5 mm), fully undo the 5 screws 32,

disconnect the upper section 29 from the lower section 31 and re-

move the sealing ring 30.

• Thoroughly rinse off the sealing ring, lower section and upper sec-

tion of the ultrafiltration cell with ultra-pure water.

Only ultra-pure water or ethanol may be used to clean the ultrafiltration

cell. Using organic solvents (e.g. acetone) will damage the Plexiglass

cell!

29

31

30

32

28

33

34

35

36