3.5 Connection to Metrohm IC systems

44

Metrohm 838 Advanced Sample Processor, Installation

3.5.5 Capillary connections for sample dilutions

A

B

C

D

E

Pump

Waste

H

F

G

Water

801

Figure 33 Capilliary connections for sample dilutions

A

Capillary (6.1803.120)

IØ = 0.97 mm, L = 3.0 m

E

Capillary (6.1803.060)

IØ = 0.97 mm, L = 0.2 m

B

Capillary (6.1803.070)
IØ = 0.97 mm, L = 0.4 m

F Capillary (6.1803.080)

IØ = 0.97 mm, L = 1.0 m

C

Capillary (6.1803.070) *
IØ = 0.97 mm, L = 0.4 m

G

Capillary (6.1803.080)
IØ = 0.97 mm, L = 1.0 m

D

Capillary (6.1803.080)

IØ = 0.97 mm, L = 1.0 m

H

Injection loop

* The

capillary C serves to aspirate. It must extend down to the bottom of the mixing container

down to the lowest point of the inclined container bottom.

1

800 Dosino

with 10 ml dosing unit, filled with water

5

PE bottle (6.1608.080)

as rinsing vessel,
on sample rack (6.2041.440)

2

Switchover valve on the 838
Note tubing connection

6 Peristaltic pump on 838 with

pump tubing LFL grey/grey
(6.1826.140)

3

Mixing container (6.1456.210)
with 801 magnetic stirrer and titration
head 6.1414.070

7 Injection valve in 820 Separation

Centre
Note tubing connection

4

Sample vessels (6.2743.050)
with PP stopper (6.2743.070),
on sample rack (6.2041.440)

8 IC column

838 Advanced
Sample Processor

820 Separation Centre

1

2

3

4

5

6

7

8