3.5 Connection to Metrohm IC systems

Metrohm 838 Advanced Sample Processor, Installation

47

3.5.8 MIC-3 modular anionic and cationic system with chemical suppres-

sion

This system requires the model 2.838.0010 of the 838 Advanced

Sample Processor.

The combined anionic/cationic IC system can operate either with

chemical suppression or with electronic suppression. The 838 Ad-

vanced Sample Processor is connected to the System MIC-3 with two

819 IC detectors, 820 IC Separation Centre (2 injectors), two 818 IC

pumps, 833 IC Suppressor Module and 830 IC Interface in accordance

with Figure 36 using the 6.2125.120 remote adapter, the 6.2125.090

remote cable and the 6.2134.100 RS232 cable.

838

819

819

830

820

818

PC

6.2134.100

6.2125.090

6.2134.100

6.2141.190

833 MSM

6.2125.120

818

6.2125.090

6.2134.090

6.2143.210

6.2115.070

6.2128.180

6.2134.040

6.2141.120

6.2134.130

Figure 36 MIC-3 interconnection with modular anionic and cationic system

with chemical suppression

838-specific settings in the «IC Net program 2.3»

Configuration of additional remote lines for the 838 Advanced Sample Processor:

Fill A 2

***0*010***

Inject A 2 ***1*000***
Reset 2

***0*000***

ZERO 2

***0*011***

Program for the 838 Advanced Sample Processor:

001 Move

sample

002 Lift

work

003 Ctrl

ZERO

1

004 Ctrl FILL A 1
005 Ctrl STEP MSM 833
006 Pump 120 s
007 Ctrl INJECT A 1
008 Move

Sample+1

009 Lift

work

− Move needle to sample position
− Move lift with needle to working height
− Trigger autozero on 819/1 IC Detector
− Switch over injection valve 1 on 820 to "Fill"
− Trigger step in the Suppressor Unit
− Fill sample loop for 120 s with sample
− Switch over injection valve 1 on 820 to "Inject"
− Move needle to next sample
− Move lift with needle to working height