3.3 Fitting the accessories

838 Advanced Sample Processor, Installation

27

3.3.11 Connection of the ultrafiltration cell

60 PEEK pressure screw (6.2744.010)

56 Pump tubing (6.1826.070)

61

Pump tubing (6.1826.030)

62 PEEK coupling (6.2744.030)

63 PTFE capillary (6.1803.050)

64

PVDF pressure screw (6.2744.000)

65 PTFE capillary (6.1803.060)

66

PTFE capillary (6.1803.080)

67 PEEK capillary (6.1831.060)

29

Upper section of the filtration cell

31

Lower section of the filtration cell

Figure 23 Connecting the

ultrafiltration cell


Use only the specified

6.2744.000 PVDF pressure screws to con-

nect to the ultrafiltration cell. If the 6.2744.010 PEEK pressure screws

are used, this may result in stress cracking on the ultrafiltration cell!

1 Pump tubing - filtration cell connection

• Fit a PEEK pressure screw 60 to one end of the PTFE capillary 65.
• Screw a coupling 62 onto this pressure screw.
• Fit a PVDF pressure screw 64 onto the other end of the PTFE capil-

lary.

• Fit this end of the PTFE capillary with the PVDF pressure screw to

the inlet opening 35 on the underside of the ultrafiltration cell (see

Figure 9).

• Fit the coupling 62 onto the outlet end of the pump tubing 56 (see

Figure 10).

2 Filtration cell - waste tubing connection

• Fit a PVDF pressure screw 64 to one end of the PTFE capillary 66.
• Fit this end of the PTFE capillary with the PVDF pressure screw to the

outlet opening 36 on the underside of the ultrafiltration cell (see Fig-

61

63

62

64

65

66

60

67

56

64

29

31

60

62

56