Advertising
background image

3.3 Fitting the accessories

32

Metrohm 838 Advanced Sample Processor, Installation

Use only the specified

6.2744.000 PVDF pressure screws to con-

nect to the dialysis cell. If the 6.2744.010 PEEK pressure screws are

used, this could cause stress cracking on the dialysis cell!

4 Connecting the injection valve

Screw the free end of the 6.1831.050 PEEK capillary 76 (L =

40 cm) with a 6.2744.010 PEEK pressure screw to connec-

tion "1" of the injection valve A 71.

Screw the 6.1831.060 PEEK capillary 77 (L =1 m) using a

6.2744.010 PEEK pressure screw to connection "2" of the in-

jection valve A 71 and guide the other end of the capillary

through one of the openings at the front or at the side of the

820 Separation Centre to the outside. Attach the other end of

the PEEK capillary 77 with a 6.2744.010 PEEK pressure

screw to the PEEK coupling on the aspiration end of the or-

ange-yellow pump tubing of channel 2 of the 833 Pump

Unit (see Chapter 3.3.9 and Figure 22).

5 Connecting the waste container

Attach the 6.1803.040 PTFE capillary 78 (L =1 m) with a

6.2744.010 PEEK pressure screw to the 6.2744.034 PEEK

coupling on the delivery end of the orange-yellow pump tub-

ing of channel 2 of the 833 Pump Unit (see Chapter 3.3.9

and Figure 22). Insert the other end of the capillary into an

adequately large waste container and secure it there.