4.2 Care and maintenance

62

Metrohm 838 Advanced Sample Processor, Service, maintenance, faults

Proceed as follows in order to exchange pump tubing:

1 Removing the old pump tubing

• Press the contact pressure lever 49 on the tubing cartridge

down fully.

• Detach the tubing cartridge 48 from the retaining bracket 50

by pushing the snap-action lever 51 in and disengage the

tubing cartridge from the retaining cam 53 on the 838 Ad-

vanced Sample Processor.

• Remove the old pump tubing.

2 Fitting the new pump tubing

• Insert the new pump tubing 56 (6.1826.xxx) into the tubing

cartridge. The stopper 57 must engage in the corresponding

fixture on the upper side of the tubing cartridge.

• Engage the tubing cartridge in the retaining cam 53 and push

it at the rear to the front until the snap-action lever 51 en-

gages on the retaining bracket 50. Ensure that the pump tub-

ing is not kinked when doing this.

3 Setting the contact pressure

• Push the pressure lever 49 upwards until the solution is just

being aspirated. Then push the pressure lever 1 further detent

position up to achieve an optimum contact pressure.

Pump tubings made of PVC or PP may not be used to rinse with solu-

tions containing acetone. Use different pump tubings or use another

pump for rinsing in this case.