3.5 Connection to Metrohm IC systems

50

Metrohm 838 Advanced Sample Processor, Installation

Dialysis.smt

Entry of the program for the 838 Advanced Sample Processor

Dialysis program:

001 Pump

on

002 Ctrl FILL A 1
003 Ctrl Pump 833 on
004 Wait

30

s

005 Ctrl INJECT A 1
006 Ctrl Pump 833 off
007 Pump

off

008 Wait

600

009 Move

Sample+1

010 Lift

work

011 Pump

on

012 Ctrl Pump 833 on
013 Wait

120

014 Ctrl Pump 833 off
015 Wait

480

s

016 Ctrl FILL B/STEP 1
017 Wait

120

018 Ctrl

ZERO

1

019 Pump

off

− Pump sample solution
− Switch over injection valve in 820 to "Fill"
− Pump acceptor solution
− Transfer time (must be determined)
− Switch over injection valve in 820 to "Inject"
− Stop pumping acceptor solution
− Stop pumping sample solution
− Waiting time for synchronisation
− Approach next sample
− Move lift with needle to working height
− Pump sample solution
− Pump acceptor solution
− Rinsing time
− Stop pumping acceptor solution, start of dialysis
− Dialysis time (120 s + 480 s = 10 min.)
− Trigger step of suppressor unit
− Wait until baseline is stable
− Trigger autozero on 819/1 IC Detector
− Stop pumping sample solution

Start-Dialysis.smt

Remove the data recorder by clicking on the recorder icon with

the right mouse button an selecting 'unlink'.

Start program:

001 Move

sample

002 Lift

work

003 Ctrl Pump 833 on
004 Pump

on

005 Wait

120

s

006 Ctrl Pump 833 off
007 Wait

480

s

008 Ctrl FILL B/STEP 1
009 Wait

180

s

010 Ctrl

ZERO

1

011 Pump

off

− Move needle to sample position
− Move lift with needle to working height
− Pump acceptor solution
− Pump sample solution
− Rinsing time
− Stop pumping acceptor solution, start of dialysis
− Dialysis time (780 s+ 120 s = 10 min)
− Trigger step of suppressor unit
− Wait until baseline is stable
− Trigger autozero on 819/1 IC Detector
− Switch over injection valve the 820 to "Inject"