3.3 Fitting the accessories

Metrohm 838 Advanced Sample Processor, Installation

33

Conditioning the dialysis system

Before the first analysis, the dialysis cell with the fitted dialysis mem-

brane and all tubing connections must be rinsed with ultra-pure water.

Please proceed as follows:

1 Settings on the 820 IC Separation Centre

In the «IC Net» software, switch injection valve A in the 820 IC

Separation Centre to position "FILL".

2 Commissioning the peristaltic pumps

Immerse the two aspiratory tubings 68 and 72 (see Figure 24)

in the acceptor solution (degassed ultra-pure water).

Switch on the 838 Advanced Sample Processor and the 833

Pump Unit with their power switches.

Switch on the peristaltic pumps of both instruments in «IC

Net» in the corresponding 'Manual Control' dialog box. Please

refer to the «IC Net» software Online Help documentation for

details of operation.

Set the contact pressure on both tubing cartridges: Push the

pressure lever 49 (see Chapter 1.3.5) up until the solutions

are just being aspirated. Then push the pressure lever 1 fur-

ther detent position upwards to achieve an optimum contact

pressure.

Flush the dialysis system for approx. 10 minutes with ultra-

pure water and check whether solution is emerging uniformly

on the two inlet capillaries leading into the waste container.

3 Conditioning the dialysis membrane

Flush the dialysis system for approx. 20 minutes with ultra-

pure water. Check whether solution is emerging uniformly on

the two inlet capillaries leading into the waste container.

Check all tubings from the supply vessels through the tubing

cartridges and the dialysis cell through to the waste vessels

for the presence of emerging fluid. If fluid is escaping at any

point, the corresponding connection must be tightened or ex-

changed.

If air bubbles remain in the dialysis cell, unscrew the PEEK

capillary 76 (acceptor solution) and PTFE capillary 73 (sam-

ple) from the outlets 44 resp. 47 of the dialysis cell and wait

until the air bubbles have disappeared. Then screw the tub-

ings back onto the dialysis cell.