Advertising
background image

3.3 Fitting the accessories

Metrohm 838 Advanced Sample Processor, Installation

31

Use only the specified

6.2744.000 PVDF pressure screws for con-

nection to the dialysis cell. If the 6.2744.010 PEEK pressure screws

are used, this could cause stress cracking on the dialysis cell!

4 Connecting the waste container

Attach the 6.1803.040 PTFE capillary 74 (L =1 m) with a

6.2744.010 PEEK pressure screw to the 6.2744.034 PEEK

coupling on the delivery end of the black-black pump tubing

of channel 2 of the peristaltic pump of the 838 Advanced

Sample Processor (see Chapter 3.3.12 and Figure 22). In-

sert the other end of the capillary into an adequately large

waste container and secure it there.

Connections for acceptor solution

1 Fitting the pump tubing to the 833 Pump Unit

Fit the two pump tubings 6.1826.030 (orange-yellow stop-

pers) as described in Chapter 3.3.9, to the 833 Pump Unit.

2 Connecting the aspiratory capillary for acceptor so-

lution

Attach one end of the 6.1803.040 aspiratory capillary 68

(L =1 m) with a 6.2744.010 PEEK pressure screw to the

PEEK coupling on the aspiration end of the orange-yellow

pump tubing of channel 1 of the 833 Pump Unit (see Chap-

ter 3.3.9 and Figure 22).

Insert the other end of the aspiratory capillary 68 into the ves-

sel with acceptor solution and secure it there. Ultra-pure

water which has been degassed beforehand (with N

2

, He or

vacuum) is normally used as the acceptor solution.

3 Connecting the dialysis cell

Attach the 6.1803.040 connection capillary 75 (L =1 m) with a

6.2744.010 PEEK pressure screw to the 6.2744.034 PEEK

coupling on the delivery end of the orange-yellow pump tub-

ing of channel 1 of the 833 Pump Unit (see Chapter 3.3.9

and Figure 22). Guide the other end of the capillary through

one of the openings at the front or on the side of the 820

Separation Centre into the column chamber and screw it onto

the inlet opening 45 of the cell upper section 38 of the dialy-

sis cell using a 6.2744.000 PVDF pressure screw 64 (see

Chapter 3.3.7 and Figure 18).

Screw the 6.1831.050 PEEK capillary 76 (L =40 cm) with a

6.2744.000 PVDF pressure screw 64 to the outlet opening 44

of the cell upper section 38 of the dialysis cell (see Chapter

3.3.7 and Figure 18).