3.5 Connection to Metrohm IC systems

Metrohm 838 Advanced Sample Processor, Installation

51

End-Dialysis.smt

End program:

001 Pump

on

002 Ctrl FILL A 1
003 Ctrl Pump 833 on
004 Wait

30

s

005 Ctrl INJECT A 1
006 Ctrl Pump 833 off
007 Pump

off

− Pump sample solution
− Switch over injection valve in 820 to "Fill"
− Pump acceptor solution
− Transfer time (must be determined)
− Switch over injection valve in 820 to "Inject"
− Stop pumping acceptor solution
− Stop pumping sample solution

The Sample Queue to be processed then contains the following entries

for instance:

System Ident.

Vial

Chrom. No.Dialysis No.

1

Start-Dialysis.smt Sample

0 1

- 1

2

Dialysis.smt Sample

1

1

1 2

3

Dialysis.smt Sample

2

2

2 3

4

Dialysis.smt Sample

3

3

3 4

5

Dialysis.smt Sample

4

4

4 5

6

Dialysis.smt Sample

5

5

5 6

7

End-Dialysis.smt Sample

6 6

6 -

This example shows processing of a sample series of 6 samples. Col-

umns 'Chrom. No.' and 'Dialysis No.' serve only for information pur-

poses in this case, Please note that the sample position specified here

is not approached in the 838 program of system 'Dialysis.smt' for dialy-

sis but that this position is incremented by one, thus processing the

next sample. Consequently, all sample vessels must be located in a

closed sequence on the sample rack.