3.5 Connection to Metrohm IC systems

48

Metrohm 838 Advanced Sample Processor, Installation

010 Ctrl

ZERO

2

012 Scan

Wait1

013 Ctrl FILL A 2
015 Pump

120

s

016 Ctrl INJECT A 2

− Trigger autozero on 819/2 IC Detector
− Wait for Remote signal
− Switch over injection valve 2 on 820 to "Fill"
− Fill sample loop for 120 s with sample
− Switch over injection valve 2 on 820 to "Inject"

3.5.9 MIC-7 anionic system with chemical suppression and dialysis

This system requires the model 2.838.0110 of the 838 Advanced

Sample Processor.

The 838 Advanced Sample Processor is connected to an anionic IC

system with chemical suppression with 819 IC detector, 820 IC Separa-

tion Centre, 818 IC pump, 833 IC Liquid Handling Unit and 830 IC Inter-

face in accordance with Figure 37 using the 6.2141.190 remote cable.

and the 6.2134.090 RS232 cable.

819

838

820

818

830

PC

6.2134.100

6.2134.130

6.2115.070

6.2134.090

6.2125.090

6.2134.100

6.2141.190

833 MSM

6.2128.180

6.2134.040

6.2125.120

6.2125.090

6.2143.200

6.2134.090

833 Pump

Figure 37 MIC-7 interconnection with anionic system with chemical suppression and

dialysis

838-specific settings in the «IC Net 2.3» program

The sequences of an automated IC analysis with prior dialysis can be

programmed in two different ways:

A) A sample is first prepared by dialysis for each analysis and this

sample is then chromatographically separated.