Dilution – Metrohm 838 Advanced Sample Processor Installation Instructions User Manual

Page 48

Advertising
background image

3.5 Connection to Metrohm IC systems

42

Metrohm 838 Advanced Sample Processor, Installation

3.5.4 MIC-2 modular anionic system with chemical suppression and sam-

ple dilution

This system requires the model 2.838.0120 of the 838 Advanced

Sample Processor.

The 838 Advanced Sample Processor is connected to an anionic IC

system with chemical suppression with 819 IC detector, 820 IC Separa-

tion Centre, 818 IC pump, 833 IC Liquid Handling Unit and 830 IC Inter-

face in accordance with Figure 32 using the 6.2141.190 remote cable

and the 6.2134.100 RS232 cable.

819

838

820

818

830

PC

6.2134.100

6.2134.130

6.2115.070

6.2134.090

6.2125.090

6.2134.100

6.2141.190

833 Pump

6.2128.180

6.2134.040

800 Dosino

801

Water

MSB1

MSB2

Figure 32 MIC-2 interconnection with anionic system with

chemical suppression and sample dilution

838-specific settings in the «IC Net 2.3» program

Example program (sample dilution) for the 838 Advanced Sample Processor:
Conditions:

• Dosino drive with 10 ml dosing unit connected to MSB connection 1

• Stirrer connected to MSB connection 2

• Mixing volume 5 ml (this is defined in «IC Net 2.3» in the configuration

of the 838)

• Rinsing vessel defined as special beaker at rack position 149

(6.2041.440 rack)

• Needle positions: "special" = 80 mm, "work" = 125 mm