3.5 Connection to Metrohm IC systems

Metrohm 838 Advanced Sample Processor, Installation

41

3.5.3 MIC-2 modular anionic system with chemical suppression

This system requires the model 2.838.0010 of the 838 Advanced

Sample Processor.

The 838 Advanced Sample Processor is connected to an anionic IC

system with chemical suppression with 819 IC detector, 820 IC Separa-

tion Centre (2.820.0230), 818 IC pump, 833 IC Liquid Handling Unit and

830 IC Interface in accordance with Figure 31 using the 6.2141.190 re-

mote cable and the 6.2134.100 RS232 cable.

819

838

820

818

830

PC

6.2134.100

6.2134.130

6.2115.070

6.2134.090

6.2125.090

6.2134.100

6.2141.190

833 Pump

6.2128.180

6.2134.040

Figure 31 MIC-2 interconnection with anionic system with

chemical suppression

838-specific settings in the «IC Net 2.3» program

Program for the 838 Advanced Sample Processor:

001 Ctrl

INIT

819

002 Move

sample

003 Lift

work

004 Ctrl

ZERO

1

005 Ctrl FILL A 1
006 Pump 240 s
007 Ctrl INJECT A 1
008 Start

DA

− Initialise remote lines on 819
− Move needle to sample position
− Move lift with needle to working height
− Trigger autozero on 819/1 IC Detector
− Switch over injection valve on 820 "Fill"
− Fill sample loop for 240 s with sample
− Switch over injection valve on 820 to "Inject"
− Start data recording