6.1 Filtration

Metrohm 838 Advanced Sample Processor, Annex

69

Test solutions:

silica particles in water

Pore size

of filter membrane 1

Effect

0.5 %, 5 μm

0.15 μm

no breakthrough

0.5 %, 5 μm

3 μm

no breakthrough

0.5 %, 5 μm

8 μm

no breakthrough

0.5 %, 5 μm

10 μm

breakthrough

2

0.5 %, 5 μm

12 μm

no breakthrough

0.5 %, 1.5 μm

0.15 μm

no breakthrough

0.5 %, 1.5 μm

3 μm

breakthrough

(1)

Nominal pore size according to information supplied by manufacturer.

(2)

All the membranes came from a single manufacturer except this one.

Table 7 Selection of the filtration membrane

Please also note that, because of their lack of thickness, the retention

capability of filter membranes may be lower than that of thicker filters

with the same nominal pore size. This should be taken into account

when choosing a suitable filter membrane.