5 GLP validation – diagnosis

5.1 Validation / GLP

GLP (Good Laboratory Practice) requires, among other things, that the

precision and correctness of analytical instruments is checked at regu-

lar intervals by using SOPs (Standard Operating Procedures).

As this instrument is not a measuring instrument as such, we recom-

mend that the Sample Processor is included in an analytical system

and is validated as part of the all-embracing validation of the whole

analytical system.

Checking the electronic and mechanical assemblies of Metrohm in-

struments can and should be undertaken within the framework of regu-

lar servicing by Metrohm technicians. All Metrohm instruments are

equipped with start-up check routines which check that the relevant as-

semblies are functioning perfectly when the instrument is switched on. If

no error message appears it can be assumed that the instrument is

functioning properly. Metrohm also supplies its instruments with built-in

diagnosis programs which allow the service technicians to check the

functioning of particular assemblies should faults or malfunctions occur

and to localize them. Diagnosis programs can also be incorporated in a

validation method.