3.3 Fitting the accessories

30

Metrohm 838 Advanced Sample Processor, Installation

Proceed as follows in order to connect the dialysis cell 37 to the IC sys-

tem:

Capillaries provided with new connectors must feature a flawless, flat

cut edge. In order to ensure this, it is best to use the

6.2621.080

cappillary cutter..

Connections for sample solution

1 Fitting the pump tubing to the 838 Advanced Sam-

ple Processor

Fit the two 6.1826.040 pump tubings (black-black stop-

pers), as described in Chapter 3.3.8, to the peristaltic pump

of the 838 Advanced Sample Processor.

2 Connecting the aspiratory capillary for sample

Attach one end of the 6.1803.040 aspiratory capillary 68 (L

=1 m) with a 6.2744.010 PEEK pressure screw to the PEEK

coupling on the aspiration end of the black-black pump tub-

ing of channel 1 of the 838 Advanced Sample Processor

(see Chapter 3.3.12 and Figure 22).

Connect the other end of the aspiratory capillary 68 to the

PEEK needle 22 (see Chapter 3.3.4 and Figure 14) and fix the

capillary in position in the guide chain of the 838 Advanced

Sample Processor.

3 Connecting the dialysis cell

Attach the 6.1803.040 PTFE capillary 72 (L =1 m) with a

6.2744.010 PEEK pressure screw to the 6.2744.034 PEEK

coupling on the delivery end of the black-black pump tubing

of channel 1 of the peristaltic pump of the 838 Advanced

Sample Processor (see Chapter 3.3.12 and Figure 22).

Guide the other end of the capillary through one of the open-

ings at the front or on the side of the 820 Separation Centre

into the column chamber and screw it on to the inlet opening

46 of the cell lower section 41 of the dialysis cell using a

6.2744.000 PVDF pressure screw 64 (see Chapter 3.3.7 and

Figure 18).

Screw the 6.1803.040 PTFE capillary 73 (L =1 m) with a

6.2744.000 PVDF pressure screw 64 to the outlet opening 47

of the cell lower section 41 of the dialysis cell (see Chapter

3.3.7 and Figure 18) and guide it through one of the openings

at the front or at the side of the 820 Separation Centre out of

the column chamber. Attach the other end of the PTFE capil-

lary 73 with a 6.2744.010 PEEK pressure screw 64 to the

PEEK coupling at the aspiration end of the black-black

pump tubing of channel 2 of the peristaltic pump of the 838

Advanced Sample Processor (see Chapter 3.3.12 and

Figure 22).