3.3 Fitting the accessories
838 Advanced Sample Processor, Installation
19
ing solution at a special beaker position on the sample rack.
• Use the IC NET software to control the 838 Advanced Sample
Processor directly. Choose the corresponding system and open the
Manual control dialog box. You can run the individual functions of
the Sampler Processor under Autosampler. Please refer to the IC
NET software Online Help documentation for details of operation.
• Move the beaker with the rinsing solution beneath the needle with
MOVE.
• You can lower the needle into the rinsing position with the LIFT
function.
• Switch on the peristaltic pump with PUMP, and then all connected
lines are rinsed and any air is removed.
• After allowing an adequate time (e.g. 5 minutes), switch the pump
back off again with PUMP and move the needle back to home posi-
tion.