15 connecting pumps, 16 connector dosing drives and stirrers, Connecting pumps – Metrohm 838 Advanced Sample Processor Installation Instructions User Manual

Page 41: Connector dosing drives and stirrers

Advertising
background image

3.3 Fitting the accessories

Metrohm 838 Advanced Sample Processor, Installation

35

3.3.15 Connecting pumps

Two external pumps can be connected and controlled on the 838 Ad-

vanced Sample Processor.

Ext.

Pump 2

The pump connections (for 3-pin M8 connectors) on the rear side of
the tower of the Advanced Sample Processor supply a voltage of 16 V
and may be loaded with max. 600 mA.

The appropriate pump models are:
• Metrohm 823 Membrane Pump Unit (membrane pump, suitable for

aqueous media without precipitation)

• Metrohm 772 Pump Unit (peristaltic pump, suitable for organic sol-

vents and aqueous solutions containing precipitation)

3.3.16 Connector dosing drives and stirrers

Metrohm Sample Processors feature various connections to which stir-

rers or dosing drives can be connected.

Stirrers

The 722/802 Rod stirrer and the 741 Magnetic stirrers feature a

DIN connector and are connected on the rear panel of the tower.

As is the case on other Metrohm instruments, the three MSB sockets

(MSB = Metrohm Serial Bus) can be used in diverse ways. Connect the

801 Stirrer at this point.

Dosing drives
Up to three dosing drives of Type 700/800 Dosino or 685/805 Dosi-

mat can be connected directly to the MSB sockets of a Sample Proc-

essor and operated. A 800 Dosino or 805 Dosimat can also be

connected via the MSB output socket of another instrument con-

nected to the MSB, such as the 801 Stirrer in a so-called 'daisy chain'. It

must be noted that only one single instrument of a specific type may

occur in such a chain, e.g. one 801 Stirrer and one 800 Dosino.
The 700 Dosino and the 685 Dosimat must be connected directly to

an MSB socket of the Sample Processor.

R S 2 3 2

M S B 1

M S B 3

M S B 2

R e m o te

Figure 27 MSB connections

A connected dosing drive is detected automatically by the instrument
only when switching on.

Dosing drive and

stirrer connections

MSB1…3

Advertising