3.3.7 Fitting the dialysis cell

Before fitting the dialysis cell, you must first insert a dialysis membrane.

Figure 18 Fitting the dialysis cell

37

Dialysis cell (6.2729.100)

43

Slot for engaging the dialysis

cell in the cell holder

38

Upper section of the dialysis

cell

44

Outlet for acceptor solution

39 Sealing ring (E.301.0111)

45

Inlet for acceptor solution

40

Dialysis membrane

(e.g. 6.2714.010)

46 Inlet for sample solution

41 Lower section of the dialysis

cell

47 Outlet for sample solution

42 Screws (V.022.6030) including

washer (4.754.4090)


Proceed as follows:

1 Preparing the dialysis cell

Remove the 6.2729.100 dialysis cell 37 from its packaging

and remove the four screwed-on 6.2744.060 blind stoppers.

Using the 6.2621.070 Allen key, fully undo the 5 screws 42,

disconnect the upper section 38 from the lower section 41

and remove the sealing ring 39.

2 Cleaning the dialysis cell

Thoroughly rinse the sealing ring 39, upper section 38 and

lower section 41 of the dialysis cell with ultra-pure water and

dry them with N

2

or with a lint-free cloth.

37

38

39

40

41

42

44

43

45

47

46

43