Counterpart User Manual - Configuration

25

DFSENS

4OUT
2OUT
8OUT
16OUT
32OUT
64OUT
128OUT

Digital filter cutout sensitivity. Specifies the number of consecutive readings that must fall
outside the filter threshold (DFTHRH parameter) before digital filtering is suspended.

DFTHRH

NONE
2D
5D
10D
20D
50D
100D
200D
250D

Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When a
specified number of consecutive scale readings (DFSENS parameter) fall outside of this
threshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.

SMPRAT

15 HZ
30 Hz
60 Hz
7.5 Hz

Sample rate. Selects measurement rate, in samples per second, of the analog-to-digital
converter. Lower sample rate values provide greater signal noise immunity.

PWRUPM

GO
DELAY

Power up mode. In GO mode, the indicator goes into operation immediately after a brief
power up display test.

In DELAY mode, the indicator performs a power up display test, then enters a 30-second
warm up period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset
and the warm up period repeated.

TAREFN

BOTH
NOTARE
PBTARE
KEYED

Enables or disables push-button and keyed tares. Possible values are:

BOTH: Both push-button and keyed tares are enabled
NOTARE: No tare allowed (gross mode only)
PBTARE: Push-button tares enabled
KEYED: Keyed tare enabled

ACCUM

OFF
ON

Accumulator. Specifies whether the scale accumulator is enabled. If enabled, accumulation
occurs whenever a print operation is performed. Scale must return to zero to re-arm a new
print.

THRESH

0

Enter a value to be used as divisions. Zero threshold allows you to select a threshold or
reset point where automatic printing functions reset themselves to be retriggered. When a
value is entered, anything that would normally require the scale to reach zero before it
happens/rearms (except for LFT parameters) now only needs to go below this value and
then above it again.

CALIBR

WZERO
WVAL
WSPAN
WLIN
REZERO

See Level 4 submenu descriptions in Table 3-7 on page 37.

SETUP Menu

Parameter

Choices

Description

Table 3-3. Setup Menu Parameters (Continued)