7 calibration menu, Counterpart, Figure 3-13. calibration menu calibr menu – Rice Lake Counterpart Configurable Counting Indicator - User Manual - Version 2.3 User Manual

Page 51: Table 3-9. calibration menu parameters

Advertising
background image

Counterpart User Manual - Configuration

45

3.7

Calibration Menu

See Section 4.0 on page 49 for calibration procedures. The Calibration menu can be protected by assigning a
password in the Feature menu.

WZERO

Calibrating

Calibrating

Display and edi

Press the calibrate

softkey to calibrate

zero.

t

test weight value

WVAL

WSPAN

Press calibrate

softkey to remove

offset from zero from

zero and span

calibrations

REZERO

WLI N

POINT 1

POINT 2

POINT 3

POINT 4

POINT 5

Same as POINT 1

WGT 1

CAL1

Calibrating

Press the calibrate

softkey to calibrate

span.

XXXXXXX

ID

DISPLAY

TEST

SETUP

AUDIT

CALIBR

Press Menu

to cancel

Press Menu to

cancel

Previous A/D raw

counts are shown

Previous A/D raw

counts are shown

SCALE 1

SCALE 2

SETPTS

Note


The

Counterpart

requires the WZERO and WSPAN points to be calibrated. The linearity points are

optional, but must NOT duplicate zero or span. During calibration, the TARE key acts as a data entry

confirmation key. It also acts as an EXECUTE key, and accepts the value if calibration was successful.

Figure 3-13. Calibration Menu

CALIBR Menu

Parameter

Choices

Description

Level 2 submenus

WZERO

Press the calibrate softkey and the A/D raw counts will be displayed.

Press calibrate softkey again to calibrate zero, or press MENU to cancel. “CALIBRATING” will
be appear prior to automatically moving to WVAL.

WVAL

Press calibrate softkey to display and edit the test weight value. Press calibrate softkey again
to move to WSPAN.

WSPAN

Press calibrate softkey and the A/D raw counts will be displayed. Press calibrate softkey to
calibrate the span or press MENU to cancel. “CALIBRATING” will be appear prior to
automatically moving to WLIN.

WLIN

POINT 1 —
POINT 5

WGT 1 allows you to display and edit the test weight value. Press calibrate softkey to edit the
value. CAL1 allows you to calibrate and display the raw A/D value. Press TARE to perform
calibration. Press calibrate softkey to will move to the next calibration point.

Note

The linearity points are optional, but must NOT duplicate zero or span.
They must be between zero and span.

REZERO

Press calibrate softkey to remove an offset value from the zero and span calibrations (e.g. if
hooks or chains are being used during calibration).

Note

Use this parameter only after WZERO and WSPAN have been set. See
Section 4.1 on page 49
for more information about using this parameter.

Table 3-9. Calibration Menu Parameters