89

9.11 Updating Firmware

To update firmware of the

Counterpart

, you must have Revolution installed and a .hex file on your computer.

Visit

www.ricelake.com

to download this free configuration software and the latest .hex file.

Note

If the hex file is the same version as is currently in the indicator, the firmware update will not reset the
configuration. This is helpful if the firmware becomes corrupt and you want to reload the same firmware.
However, it is always recommended that configuration be backed up in Revolution to avoid any data loss.

1. Unplug power to the

Counterpart

and remove top from the base of the unit.

2. Wire the serial cable to the

Counterpart

’s Port 1. Refer to Table 9-7 for cable connection settings.

Table 9-7. Cable connection settings

DB9 Cable

Connector (J5)

Pin/Color

Signal

In/Out

Description

Pin

Description

In/Out

1/Brown

DCD

In

Data Carrier Detect

NC

NA

NA

2/Red

RxD

In

Receive Data

3

Tx

Out

3/Orange

TxD

Out

Transmit Data

4

Rx

In

4/Yellow

DTR

Out

Data Terminal Ready

5

DTR

In

5/Green

GND

--------

Ground

1 or 2

Gnd

--------

6/Blue

DSR

In

Data Set Ready

NC

NA

NA

7/Purple

RTS

Out

Request to Send

6

RTS

In

8/Gray

CTS

In

Clear to Send

NC

NA

NA

9/Black

RI

In

Ring Indicator

NC

NA

NA

3. Install jumpers on JP5 and JP6. See Figure 2-4 on page 9.

4. Plug in power to the

Counterpart

and press the indicator’s

Power

button.

5. With Revolution open, begin a new configuration file for the

Counterpart

.

6. Click

Update

Counterpart

Firmware

.

Figure 9-4. Revolution Screen.

7. The Rice Lake Updater screen appears. Specify the COM port the

Counterpart

is connected to, and click

the ellipses (...) to browse to and select the desired .hex file.