Counterpart User Manual - Calibration

51

4.5

Revolution

®

Calibration

To calibrate the indicator using Revolution, the
indicator EDP port must be connected to a PC running
the Revolution configuration utility.

1. Place the indicator in config mode (display

reads

SCALE

, see

Figure 3-6 on page 23) and

remove all weight from the scale platform.

2. From Revolution, select

File

»

New

.

The Select Indicator dialog box appears.

3. Select

Counterpart

and click

OK

.

4. From the Communications menu, select

Connect

.

5. From the left pane, expand the Scale selection

and click the

Scale

button.

Figure 4-2. Scale Button

6. From the Tools menu, select

Calibration Wizard

.

7. Click

NEXT

to begin the Calibration Wizard.

8. Select whether you want to perform a

standard calibration or a standard with
multi-point linearization and click

NEXT

.

9. In the text box, enter the test weight value be

used for span calibration.

10. Select the checkbox if you are using chains or

hooks during the calibration, then click

NEXT.

11. Remove all weight from the scale and click

Click to Calibrate Zero

to begin zero calibration.

If your test weights require hooks or chains,
place them on the scale for zero calibration.

12. When zero calibration is complete, the

Calibration Wizard prompts you to place test
weights on the scale. Place the test weights on
the scale, then click

Click to Calibrate Span

.

13. If you chose to perform linear calibration, the

Calibration Wizard now displays prompts
(1–5). Enter the weight value for Linear Point
#1, place test weights on scale and click

GO

.

Repeat for additional linearization points,
then click

NEXT

.

14. The new and old calibration settings are

displayed. To accept the new values, click

Finish

. To exit and restore the old values, click

Cancel

.

4.6

More About Calibration

The following topics provide additional information
about compensating for environmental factors
(Section 4.6.1) and diagnostic information for
determining expected zero and span coefficients.

4.6.1

Adjusting Final Calibration

Calibration may be affected by environmental factors
including wind, vibration, and angular loading. For
example, if the scale is calibrated with 1000 lb, a
s t r a i n t e s t m a y d e t e r m i n e t h a t a t 2 0 0 0 l b t h e
calibration is 3 lb high. In this case, final calibration
can be adjusted by tweaking WVAL to 998.5 lb. This
adjustment provides a linear correction of 1.5 lb per
1000 lb.

To adjust the final calibration, return to the

WVAL

prompt and press

TARE(Enter)

to show the test weight

value. Press or to adjust calibration up or down.
Press

Calibrate softkey

to save the value, then press to

return to the CALIBR menu.