4.2

Five-point linearization

Five-point linearization (using the WLIN parameter)
provides increased scale accuracy by calibrating the
indicator at up to five additional points between the
zero and span calibrations.

Linearization is optional: if you choose not to perform
l i n e a r i z a t i o n , s k i p t h e W L I N p a r a m e t e r ; i f
linearization values have previously been entered,
these values are reset to zero during calibration.
WZERO and WSPAN must be calibrated before
adding linearization points; WLIN values must be less
than the WSPAN value and must not duplicate
WZERO or WSPAN.

To perform linearization, follow the procedure below:

With

WLIN

displayed, Press

down

to go to the first

linearization point (

POINT 1

). Press

down

again to

show the weight value prompt (

WGT 1

), then

down

once more to show the weight value. Place test
weights on the scale, then use the numeric keypad to
enter the actual test weight value. Press

enter

to save

the value and move to the calibration (

CAL 1

) prompt.

Press

down

to show the current calibration value, then

press the

Calibrate

softkey to calibrate the linearization

point. When complete, the A/D count for the linear
calibration is displayed. Press

enter

again to save the

calibration value and go to the next prompt (

POINT 2

).

Repeat for up to five linearization points. To exit the
linearization parameters, press the

up

key to return to

WLIN.

4.3

Rezero

The optional rezero function is used to remove a
calibration offset when hooks or chains are used to
hang the test weights.

NOTE:

The rezero function cannot be used with

five-point linear calibration.

If no other apparatus was used to hang the test weights
during calibration, remove the test weights and press

up

to return to the CALIBR submenu.

If hooks or chains were used during calibration,
remove these and the test weights from the scale. With
all weight removed, go to the REZERO parameter,
then press

down

to show the current zero value. Press

the

Calibrate

softkey to adjust the zero and span

calibration values. Press

enter

or

up

to return to the

CALIBR submenu.

Press

up

to return to the SCALES menu, or press the

Save and Exit

softkey to exit setup mode.

4.4

EDP Command Calibration

To calibrate the indicator using EDP commands, the
indicator EDP port must be connected to a terminal or
personal computer. See Section 2.7.1 on page 9 for
EDP port pin assignments.
Once the indicator is connected to the sending device,
do the following:

1. Place the indicator in config mode (display

must read

SCALE

- see

Figure 3-6 on page 23)

a n d r e m o v e a l l w e i g h t f r o m t h e s c a l e
platform. If your test weights require hooks or
chains, place the hooks or chains on the scale
for zero calibration.

2. Send the SC.WZERO#n command to

c a l i b r a t e z e r o . T h e i n d i c a t o r d i s p l a y s

CALIBRATING

while calibration is in progress.

3. Place test weights on the scale and use the

SC.WVAL#n command to enter the test
weight value in the following format:

SC.WVAL=nnnnnn<CR>

4. Send the SC.WSPAN#n command to calibrate

span. The indicator displays

CALIBRATING

while calibration is in progress.

5. Up to five linearization points can be

c a l i b r a t e d b e t w e e n t h e z e r o a n d s p a n
c a l i b r a t i o n v a l u e s . U s e t h e f o l l o w i n g
commands to set and calibrate a single
linearization point:

SC.WLIN.V1=nnnnn<CR>
SC.WLIN.C1<CR>

The SC.WLIN.V1#n command sets the test
weight value (nnnnn) for linearization point 1.
The SC.WLIN.C1#n command calibrates the
point. Repeat using the SC.WLIN.Vx and
SC.WLIN.Cx commands as required for
additional linearization points.

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used
t o h a n g t e s t w e i g h t s , t h e n s e n d t h e
SC.REZERO#n command. The indicator
displays

CALIBRATING

while the zero and span

calibrations are adjusted.

7. Send the SC.KUPARROW#n or

SC.KEXIT#n EDP command to return to
normal mode.