Page 106 MSI-9850 RF Remote Indicator • User Guide

MSI-9850

HANDHELD RF REMOTE INDICATOR for

C

ELL

S

CALE

®

CHANNEL FILTER

1 Off

2 Low

3 High

1 Channel INACTIVE

1 Channel is ACTIVE

CALIBRATE -CHN-

SETTINGS

1 Motion Band 0d

2 Standard INDUSTRY

3 Center of Zero OFF

4 AutoZero (AZM) OFF

5 AZM Band 1d

Starts the Calibration

Procedure

3 Center of Zero OFF

3 Center of Zero ON

SET MOTION BAND (d)

ESC exits no change

ENTER saves value

0-9 replaces value

10

min = 0 max = 255

^ incs, v decs digit

SCALE CAL STANDARD

1 Industrial (unreg)

2 NIST NTEP

3 OIML R76

4 Industrial Metric

4 AutoZero (AZM) OFF

4 AutoZero (AZM) ON

AUTOZERO BAND (d)

ESC exits no change

ENTER saves value

0-9 replaces value

10

min = 0 max = 255

^ incs, v decs digit

5 Zero RUSure Msg OFF

5 Zero RUSure Msg ON

CHANNEL / CALIBRATION

--Channel Name--

--Chn-- --Index--

1 Channel is ACTIVE

2 Filter LOW

3 Calibrate Settings

4 Start Calibration

5 Zero RUSure Msg ON

Changes to the

next Channel

available

Shortcut

CHANNEL / CALIBRATION

--Channel Name--

--Chn-- --Index--

CALIBRATE 1-1

SETTINGS

CHANNEL FILTER

SET MOTION BAND (d)

SCALE CAL STANDARD

AUTOZERO BAND (d)

ESC

!

SETUP

2

DEF

1

ABC

3

GHI

5

MNO

4

JKL

7

STU

F2

CHANNEL

1

ABC

2

DEF

3

GHI

4

JKL

1

ABC

F2

CHANNEL

5

MNO

CHANNEL / CALIBRATE SETTINGS

Go to Index