Page 90 MSI-9850 RF Remote Indicator • User Guide

MSI-9850

HANDHELD RF REMOTE INDICATOR for

C

ELL

S

CALE

®

SECTION 14 – CHANNEL SETUP & CALIBRATION

CHANNEL SETUP MENU

CHANNEL FILTER

1 Off

2 Low

3 High

1 Channel INACTIVE

1 Channel is ACTIVE

CALIBRATE -CHN-

SETTINGS

1 Motion Band 0d

2 Standard INDUSTRY

3 Center of Zero OFF

4 AutoZero (AZM) OFF

5 AZM Band 1d

Starts the Calibration

Procedure

3 Center of Zero OFF

3 Center of Zero ON

SET MOTION BAND (d)

ESC exits no change

ENTER saves value

0-9 replaces value

10

min = 0 max = 255

^ incs, v decs digit

SCALE CAL STANDARD

1 Industrial (unreg)

2 NIST NTEP

3 OIML R76

4 Industrial Metric

4 AutoZero (AZM) OFF

4 AutoZero (AZM) ON

AUTOZERO BAND (d)

ESC exits no change

ENTER saves value

0-9 replaces value

10

min = 0 max = 255

^ incs, v decs digit

5 Zero RUSure Msg OFF

5 Zero RUSure Msg ON

CHANNEL / CALIBRATION

--Channel Name--

--Chn-- --Index--

1 Channel is ACTIVE

2 Filter LOW

3 Calibrate Settings

4 Start Calibration

5 Zero RUSure Msg ON

Changes to the

next Channel

available

Shortcut

CHANNEL / CALIBRATION

--Channel Name--

--Chn-- --Index--

CALIBRATE 1-1

SETTINGS

CHANNEL FILTER

SET MOTION BAND (d)

SCALE CAL STANDARD

AUTOZERO BAND (d)

ESC

!

SETUP

2

DEF

1

ABC

3

GHI

5

MNO

4

JKL

7

STU

F2

CHANNEL

1

ABC

2

DEF

3

GHI

4

JKL

1

ABC

F2

CHANNEL

5

MNO

The Channel Setup Menu provides control of Channel settings in

the CellScale. The CHANNEL / CALIBRATION menu screen

will display the current channel name (if it was provided in the

CellScale), channel number and scan list index number.

[1]

Channel is ACTIVE / INACTIVE – The Channel is operat-

ing as part of the Scan List if ACTIVE. If set INACTIVE, the

channel will not be available to any device in the Network,

and will not be displayed on the 9850.

[2]

Filter – Each A/D channel has independent filtering set by

this menu. Set “Off” for normal scales, Low or High for

scales that are subject to a lot of movement or vibration, or

for very high resolution scales. This menu is locked out in

L-F-T systems until the calibration switch is enabled at the

CellScale. Advanced settings for filtering must be performed

in the Master CellScale.

[3]

Calibrate Settings – See following submenu description.

This menu is locked out in L-F-T systems until the calibra-

tion switch is enabled at the CellScale.

[4]

Start Calibration – Starts the Calibration Procedure. This

menu is locked out in L-F-T systems until the calibration

switch is enabled at the CellScale.

CALIBRATE SETTINGS

[1]

Motion Band – The motion band determines the range of

weight variation that the indicator considers to be stable

weight. Selecting “Motion Band” calls a numeric entry

screen.

[2]

Scale Cal Standard – The submenu has 4 choices: Industrial

[1]

No limits on settings. NIST NTEP

[2]

Settings limited

to NTEP ranges, calibration is protected by the CellScale

Sealed Switch. OIML R76

[3]

Settings limited to R76

ranges, only metric units allowed. Industrial Metric

[4]

Same as 1 except only metric units allowed.

[3]

Center of Zero ON/OFF – Enables the Center of Zero

annunciator. The scale is within 1/4 d of center when the

annunciator is on.

[4]

Autozero (AZM) ON/OFF – Enables and disables Auto

Zero Maintenance (AZM).

[5]

Autozero Band – Calls a numeric entry screen to set up the

range of weight in ‘d’ that is automatically zeroed. Typical

settings for the AZM band are ±1d for most applications, ±3d

for truck scales, and ±.6d for L-F-T applications (reached

by setting the band to 0)

Go to Index