Program set point menu – Rice Lake MSI-9850 CellScale RF Digital Indicator - Operator Manual User Manual

Page 69

Advertising
background image

MSI CellScale

®

System • 9850 User Guide Page 69

MEASUREMENT SYSTEMS INTERNATIONAL

Firmware Version 5-XX for 2450 Modems

PROGRAM SET POINT MENU

1 SP

XX -SPName- OFF

2 9750 Response

OFF

3 Input Source

XX

4 Output Assign

XX

5 Delay

XXXXXms

6 Latch

OFF

7 Force On

OFF

8 Formula 9 View

1 SP

XX -SPName-

ON

1 SP

XX -SPName-

OFF

Len 0 F4 Menu

Max 8 ENT Save

Pos 1

SETPOINT XX NAME

1

VIEW FORMULA SP

XX

Set Pt Triggers If

N/G Wt >= 500.0

AND N/G Wt <=1000.0

Set Pt Resets If

N/G Wt < 500.0

OR N/G Wt > 1000.0

SET POINT DELAY ms

ESC exits no change

ENTER saves value

0-9 replaces value

1000

min = 0 max =10000

^ incs, v decs digit

OUTPUT SCAN LIST LOC

ESC exits no change

ENTER edits value

0-9 replaces value

1

min = 1 max = 32

^ incs, v decs digit

6 Latch ON

6 Latch OFF

7 Force Output ON

7 Force Output OFF

INPUT SCAN LIST LOC

ESC exits no change

ENTER edits value

0-9 replaces value

1

min = 1 max = 32

^ incs, v decs digit

To 9750 Response Menu

To Set Point Formula Menu

1

ABC

See General Text Editing

Procedure

To edit Set Point

Name

ALPHA

A

a

1

2

DEF

3

GHI

6

PQR

5

MNO

4

JKL

7

STU

8

VWX

9

@YZ

SET POINT DELAY ms

SET POINT OUTPUT LIST

INPUT SCAN LIST LOC

This Set Point Menu is reached from the procedure on

the previous page.
1) Individual Set Points are turned on (enabled) or

off (disabled) by pressing

[1]

. Disabled Set Points

can still be programmed, but will not operate until

enabled.

2) 9850 Response – This menu item sets the response

modes of the 9850. The submenus are detailed on the

next page.

3) Input Source – Input the scan channel index number

(1-32) for the scale input or math channel this set

point is compared to.

In a single Scale channel CellScale system, the input source would always
be “1” as this would be the only channel in the scan list.

4) Output Scan List Location – Input the Output List

choice for this set point. Output lists are set in the

CellScale. The output list controls the CellScale

response and has no effect on the 9850.

5) Set Point Delay ms – Set point delays are used to

eliminate spurious alarms on transient events or to

delay the set point action. Enter the desired delay in

milliseconds (1000ms = 1 second). 0 to 10000ms

range (10s). Actual resolution is 50ms. The number

will be rounded to the nearest 50ms.

6) Latch – A set point latch forces the set point output

to remain on even after the set point trip condition

is removed. Defaults to OFF.

7) Force Output – Used to test external response settings

of the set point. Forcing the output on will cause all

set point responses to operate. Be sure to force the

output back off when finished.

Caution: This function will turn on relays, warnings, alarms, etc. Be sure that
no damage will result to equipment by forcing the outputs on. This function
is automatically cleared when leaving the set point menus.

8) Formula – Use to program the set point formula. See

“Input Formula Menu”.

9) View – Shows trigger and reset points for this set

point. Used to confirm the formula. N/G indicates

the set point responds to Net or gross weight.