Calibrate general information, Enable calibration, To calibrate – Rice Lake MSI-9850 CellScale RF Digital Indicator - Operator Manual User Manual

Page 91

Advertising
background image

MSI CellScale

®

System • 9850 User Guide Page 91

MEASUREMENT SYSTEMS INTERNATIONAL

Firmware Version 5-XX for 2450 Modems

CALIBRATE GENERAL INFORMATION

The following sections are intended for qualified scale technicians. The CellScale can be calibrated either directly

using the Terminal Interface mode, or by using a 9850 or 3750CS. See the CellScale operators guide for informa-

tion on direct calibration. This section details the calibration of a CellScale

using the 9850 RF Remote Indicator.

Calibrating a CellScale with a 9850 requires that adequate test weights (at least 10% of capacity) are available.

The CellScale is capable of multi-point calibration with any combination of weights from zero to capacity.

Calibration constants are stored only in the CellScale. The 9850 serves only as a controller during calibration.

Since a single CellScale is capable of hosting up to

32 independent scales, the scale channel to be calibrated must be selected first using the CHANNEL Key. To Calibrate multiple Scale Inputs, you must complete
the first calibration. Then exit out of the Calibration Menus. Select the next scale with the CHANNEL Key. Then enable Calibration again.

ENABLE CALIBRATION

Calibration of a CellScale from the 9850 may be protected in either of two ways: 1) Legal-for-Trade systems are

protected with a sealed push button switch on the CellScale. This switch must be pressed to enable the calibra-

tion menus. 2) The 9850 also offers password locks to prevent calibration. See the Password locks section for

information on how to set passwords.
Industrial use scales are not required to have the calibration protected. However MSI recommends using the

Password locks to prevent accidental calibrations.
To Enable Calibration (Legal-for-Trade CellScales Only)

1) Expose the Calibration Seal Port by removing the seal screw on the CellScale with a Phillips Head Screw-

driver. Seal Screw location can be found in the CellScale manual or the 9300 Crane Scale Manual.

2) Insert a small non-metallic screwdriver or the wooden stem of a “Q-Tip” and press and release the switch

button in the hole. The CellScale will acknowledge the press button by flashing its LEDs.

3) Now you can proceed with calibration and calibration settings.

TO CALIBRATE

Use this Calibration procedure for calibration of the CellScale using the 9850. The CellScale supports 10 point

calibration - zero and up to 9 points of span calibration. Zero and at least 1 span point are required for a successful

calibration. The highest value test weight should be at least 10% of capacity, but 50% or more is preferable.

Calibration Procedure

1) Multichannel systems only - Using the

CHANNEL

key, step to the Scale channel you wish to calibrate.

Only A/D channels can be calibrated (1-x, 2-x, 3-x, and 4-x). Do not attempt to calibrate a math channel

(6-x).

2) Press

SETUP

followed by

CHANNEL

, or

SETUP

, “MORE”

[7]

, and select “Channel / Calibration”

[1]

.

3) Select “Start Calibration”

[4]

.

4) If this is a re calibration, press

ENTER

and jump to step 9. If this is a new installation, or you wish to

change the capacity and/or scale division size, press

CLR

(

ZERO

).

5) The screen displays “RESETTING CALIBRATION” a few seconds while the calibration memory is

purged.

6) Select a calibration unit by pressing

1

-

7

. This should be in the units the scale is most commonly used in.

However, sometimes it is desirable to use the units your test weights are cal’d in for convenience in number

entry. You can always change the units for display later.

7) Enter the capacity of the scale (in the cal unit selected) using the numeric keypad. Press

ENTER

.

8) The CellScale calculates the appropriate scale divisions based on allowed resolution and then sends them

to the 9850 as a list. The “SCALE DIVISIONS (d)” menu will vary depending on capacity and units. From

5 to 12 possible divisions will be shown. Pick the rated scale division for your scale.

Avoid excess total divisions (capacity/d = total divisions) for the type of scale you are using. Crane scales should be less than 5000 total divisions and
depending on usage sometimes less than 1000 divisions. Platform scales are rarely accurate above 10000 divisions. Lower resolution means faster settling
time, and a more stable display in vibration and motion situations.

9) “ZERO CALIBRATION” – Unload the scale. When the scale is stable press

ENTER

. There will be a short

delay while the A/D in the CellScale performs all its initializations.

Go to Index

Advertising
This manual is related to the following products: