Msi-9850, Reseting calibration – Rice Lake MSI-9850 CellScale RF Digital Indicator - Operator Manual User Manual

Page 92

Advertising
background image

Page 92 MSI-9850 RF Remote Indicator • User Guide

MSI-9850

HANDHELD RF REMOTE INDICATOR for

C

ELL

S

CALE

®

ENTER CAPACITY

ESC exits no change

ENTER saves value

0-9 replaces value

17

min = 2, max =999999

^ incs, v decs value

CHANNEL / CALIBRATION

--Channel Name--

--Chn-- --Index--

1 Channel is ACTIVE

2 Filter LOW

3 Calibrate Settings

4 Start Calibration

5 Zero RUSure Msg ON

ENTER to

Recalibrate

CLR to Reset Cal

ESC to Exit Cal

CALIBRATION UNITS

1 Pounds

2 Kilograms

3 Tons (US Short)

4 Metric Tons

5 DekaNewtons

6 Ounces

7 Grams

Changes to the

next Channel

available

SCALE DIVISIONS (d)

-Pounds-

1 1 7 0.1

2 2

8 0.05

3 5

9 0.02

4 10

0 0.01

5 0.5

F3 0.005

6 0.2

ZERO CALIBRATION

Unload scale. Press

ENTER when stable.

ESCape ends Cal.

SPAN CALIBRATION 1

Scale =

-XXXXXXX-

Cal Wt 1 =

-XXXXXXX-

0-9 New Cal Weight

When loaded & stable

press ENTER to Cal

ESCape to end Cal

SPAN CALIBRATION

-2-

Scale =

-XXXXXXX-

Cal Wt 2 =

-XXXXXXX-

0-9 New Cal Weight

When loaded & stable

press ENTER to Cal

ESCape to end Cal

RESETING

CALIBRATION

Up to 9 Span Points

Shortcut

4 CalibrationDisabled

CALIBRATION

DISABLED, CYCLE

CELLSCALE POWER

TO CONTINUE

OR

CALIBRATION UNITS

CHANNEL / CALIBRATION

--Channel Name--

--Chn-- --Index--

CALIBRATION LOCK

PRESS SEALSWITCH

ON CELLSCALE TO

CONTINUE

If Disabled (Sealed)

ESC

!

SETUP

F2

CHANNEL

1

ABC

7

STU

F2

CHANNEL

ZERO

CLR

0

ENTER

4

JKL

10) “SPAN CALIBRATION 1” – The CellScale transmits back the

capacity as a possible calibration weight (Cal Wt 1 =). Load the

scale with the first calibration weight. If your cal weight is not equal

to capacity, enter the cal weight with the keypad. When the weight

is stable press

ENTER

.

Any calibration weight is possible. MSI recommends a cal weight of at least 10% of capacity for
industrial systems. For Legal-for-trade systems MSI recommends using at least 50% of capacity,
and consider multipoint calibration.

11) “SPAN CALIBRATION 2” – Now that the CellScale has 1 span

point, it transmits back the current reading. If a single span point

is adequate, press

ESC

(

SETUP

) to end calibration. If multipoint

calibration is desired, load the next cal weight. If the displayed scale

weight agrees with the cal weight, you do not need to calibrate

the point. However if the weight displayed does not match the cal

weight, use the keypad to enter the correct weight. When the weight

is stable press

ENTER

.

12) Repeat step 11 up to a total of 9 span points. Span points can be

in any order, low to high, high to low, jumping around...whatever.

At any time, pressing the

ESC

(Setup) key will end calibration.

Even non-linear transducers can be linearized with the judicious

application of multi-point calibration.

Go to Index