5 adjusting the vertically adjustable castors, Danger – BINDER UF V 500 User Manual

Page 34

Advertising
background image

UF V (E2+E2.1) 03/2013

page 34/112

Avoid direct solar radiation on the unit. Do not place the freezer in direct vicinity of units with a

high heat emission.

Permissible ambient humidity: 70% r.H. max., non-condensing.

Installation height: max. 2000 m / 6561.7 ft above sea level.


Minimum distances:

between several units: 250 mm / 9.84 in

Wall distance, rear: 100 mm / 3.94 in (spacer is supplied, see chap. 4.2)

Wall distance, laterally, on the side without door hinge: 100 mm / 3.94 in

Wall distance, laterally, on the side with door hinge: 240 mm / 9.45 in.

Spacing above the unit: 100 mm / 3.94 in

Ventilation openings must not be blocked. Ensure a distance of at least 100 mm / 3.94 in to the ventilation

openings on the freezer’s front and rear.

To completely separate the unit from the power supply, you must disconnect the power plug.

Install the unit in a way that the power plug is easily accessible and can be easily pulled in

case of danger. To disconnect the connector plug, simultaneously pull the locking button

(Figure 23).

With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing)

several times a year. Check the condenser air filter frequently and clean it if necessary (chap. 14.4.1).
Avoid any conductive dust in the ambiance according to the unit layout complying with pollution degree 2

(IEC 61010-1).
Do not install or operate the freezer in potentially explosive areas.

DANGER

Explosion hazard.
Danger of death.

Do NOT operate the unit in potentially explosive areas.

KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the unit.


3.5 Adjusting the vertically adjustable castors

The vertically adjustable castors with locks serve to block the castors and to set up the unit at level (op-

erating position).
For transport raise the castor locks completely (moving position, live ring leftwards) to achieve maximum

ground clearance. For setting up, let them down (operating position, turn live ring rightwards, Figure 14).

Advertising
This manual is related to the following products: