BINDER UF V 500 User Manual

Page 93

Advertising
background image

UF V (E2+E2.1) 03/2013

page 93/112


Fault description

Possible cause

Required measures

Refrigerating performance (continued)

Alarm message A41 (cooling

cycle operating permanently > 72

hs.)

Door not shut tightly.

Check if doors are closed.

Frost on door gasket.

Defrost the door gasket with the

ice scraper.

Door gasket defective.

Contact BINDER service.

Door opened very frequently.

Open doors less frequently

Place of installation too warm.

Select cooler place of installation

or contact BINDER service.

Introduction of too warm or too

large amount of material.

Cool down material before intro-

ducing and / or load in smaller

portions.

Too low temperature set-point set

on the controller.

Set temperature set-point correct-

ly on the controller

Controller not adjusted, or ad-

justment interval exceeded.

Calibrate and adjust controller

Semiconductor relay defective.

Contact BINDER service.

Controller defective.

Alarm message A51

Cooling system error

Contact BINDER service.

Alarm message A52

Condenser air filter soiled.

Clean the condenser air filter

(chap. 14.4.1).

Condenser soiled.

Clean the condenser (chap.

14.4.2).

Ventilation slots are blocked.

Make sure to have free air ac-

cess to the device at the front

and bottom.

Alarm message A53

Condenser fan defective.

Turn off the freezer and contact

BINDER service.

Safety controller responds.

Limit temperature reached.

Check setting of temperature set-

point and of safety controller. If

appropriate, select suitable limit

value.

Too much external heat load.

Reduce heat load

Controller defective.

Contact BINDER service.

Safety controller defective.
Semi-conductor relay defective.

Humidity
Icing at the walls of the inner

chamber

Long time of continuous opera-

tion.

Defrost the unit (chap. 14.4.3)

Controller

No unit function

(dark display).

Power failure. Display mode

“Standby” active.

Press any key on the controller

display.

Main power switch (key switch) is

off.

Turn on the key switch (6).

No entries to controller keypad

possible.

Keyboard locking activated.

Unlock keyboard locking (chap.

8.4).

No access to menu “USER”.

User code incorrect.

Contact BINDER service.

EXIT button does not cancel the

notifying or alarm indication.

Cause of disturbance not re-

moved correctly
The EXIT button permits resetting

notifying or alarm messages for

temperature only with in a toler-

ance sector of +/- 2 °C.

Remove cause of disturbance. If

the EXIT button still does not

cancel the indication, contact

BINDER service.

Advertising