10.10 Zero-voltage relay alarm output

Collective alarm output via the zero-voltage relay alarm contact
The freezer is equipped at the rear with a zero-voltage relay output (8), which permits the transmission of

alarms to an external monitoring system in order to monitor and record the alarm signals.
The zero-voltage relay alarm output switches immediately, as soon as the red LED

lights up on the

controller display. The zero-voltage relay alarm output switches for alarm instances listed in chap. 17.2

and in case of a power failure.
If the external alarm monitor is connected via the contacts C and NO, alarm monitoring will take place

with protection against short-circuiting, i.e., if the connection between the freezer and the external alarm

monitor is interrupted, an alarm is triggered. In this case, power failure will also trigger the alarm.




Figure 55: Zero-voltage contacts circuit diagram

and pin allocation of DIN socket (8)

In case of no alarm, contact C closes with contact NO.
Closing contact C with contact NC switches the zero-voltage relay alarm output.

Maximum loading capacity of the switching contacts: 24V AC/DC – 2.0 Amp.

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 24V AC/DC – 2.0 Amp.

Do NOT connect any devices with a higher loading capacity.

The alarm message on the controller display remains displayed during transmission of an alarm via the

zero-voltage relay outputs. As soon as the cause of the alarm is rectified, or the alarm message has been

reset with the “EXIT” key, the alarm transmission via the zero-voltage relay outputs is reset together with

the alarm message on the controller display.
In case of power failure, transmission of the alarm via zero-voltage relay outputs remains active for the

duration of the power failure. Afterwards, the contact closes automatically.
Connection to an external monitoring system
To ensure short-circuit-proof alarm monitoring that will trigger the alarm when the freezer is connected to

an external alarm monitor, connect the external alarm monitoring system to the freezer via the connection

socket (8) of the zero-voltage relay output.
Connection of the BINDER GSM box for remote alarms (option)
This option allows forwarding the alarm messages, which are transmitted via the zero-voltage relay alarm

contact, to the mobile network and to receive defined messages as a text message (SMS) on mobile

phones.

The installation and operation of the BINDER GSM box are described in the manual (Art.

No. 7001-0236) which is included with the GSM box.