10.3 Safety controller temperature alarm

The selected temperature value of the safety controller was exceeded.








Figure 41: Safety controller

temperature alarm

Alarm message “

otc” immediately

LED indication immediately: LED

flashes, LED continually lit

Audible alarm: buzzer (ongoing sound) immediately

Switching the zero-voltage relay alarm output

Actions:

Check whether the outer door was open for a long time or is not closed properly. Close the door if

necessary.

Check the setting of the operating function “

SA.F” (limit temperature of the safety controller) in

operating modes HAND (chap. 8.2) or USER (chap. 8.3). The limit temperature should be approx. by

15 K above the temperature set point. If necessary, adjust the relevant value.

Check whether samples were inserted into the freezer that may release heat.

Check the ambient conditions. Protect the freezer from direct sunlight. Ensure sufficient ventilation

around the installation location to prevent any buildup of heat in the unit.

If these points do not reveal the source of the fault, it may be that the unit is faulty. Please contact

BINDER Service.