7 power failure alarm – BINDER UF V 500 User Manual

Page 67

Advertising
background image

UF V (E2+E2.1) 03/2013

page 67/112

10.7 Power failure alarm






Figure 48: Power failure alarm

No alarm message (display off)

You can turn on the display with any key. It turns off automatically after 5 seconds.

LED indication immediately: LED

flashes slowly

Audible alarm: buzzer (intermittent sound: 1 sec. on, 4 sec. off) immediately

Switching the zero-voltage relay alarm output.

WARNING: If customer should use a freezer running in non-supervised continuous opera-

tion, we strongly recommend in case of inclusion of irrecoverable specimen or samples to

split such specimen or samples and store them in at least two chambers, if this is feasible.