Start up, 1 preset factory parameters, 2 operating the automatic door mechanism – BINDER UF V 500 User Manual

Page 42: Ole_link7, Ole_link9

Advertising
background image

UF V (E2+E2.1) 03/2013

page 42/112

5.

Start up

Following transport, wait at least 8 hours until start-up.


Check that the interior of the freezer is empty. Prior to starting a new freezer

or if you do not know what the freezer was last used for, for hygiene purpos-

es you should clean and disinfect the interior (chap.14.3).
After connecting the supply lines, turn on the unit by the key switch (6) set-

ting it to position 1).

Figure 24: Key switch (position ON)

5.1 Preset factory parameters

The unit is supplied with the following preset parameters:

Temperature set point

-80 °C / -112 °F

Safety controller

Set point type “Limit”
-60 °C / -76 °F

The maximum permitted deviation from the temperature set point for

tolerance range alarm

6 K

Delay time tolerance range alarm temperature after closing the door

60 min

Password to lock / unlock the HAND operating functions

0 (no locking)

Password to access operating mode USER

1

5.2 Operating the automatic door mechanism



Opening the door:
Firmly press the one-touch pushbutton “OPEN”. You can now open

the door.
Closing the door:
Firmly press on the door for at least 2 seconds until the automatic

door mechanism is activated and closes the door tightly.

Figure 25: Pushbutton “OPEN” to

open the door

After turning on the freezer, you can open the door after 30 seconds only.
Following closing the door, you can open the door again only after approx. 30 seconds due to

necessary venting.