5.5 Behavior after turning on the unit

During the equilibration phase of 9 hours after turning on the unit, undefined temperature conditions oc-

cur within the unit. During this phase, do not place any sample materials in the unit

CAUTION

Danger of destruction of samples.

Load the freezer only after equilibration of temperature.

Following turning on the unit with the key switch (6), there is a subsequent brief startup phase in which

the display items at the edges of the controller display light up successively.
After a few seconds, the display shows the current interior temperature of the freezer. The lit LED at the

right of the display shows that the temperature is displayed in °C.








Figure 29: Normal display of controller after turning on the

freezer, showing the current temperature value: 15 °C (ex-

ample)


Set the controller to the desired temperature set point (chap. 6) used to operate the freezer.

As long as there is a difference between the actual and set value shown in the display, the

intended operation of the unit will not be ensured.