BINDER UF V 500 User Manual

Page 76

Advertising
background image

UF V (E2+E2.1) 03/2013

page 76/112


Low battery voltage during CO

2

emergency cooling with battery operation






Figure 59: Alarm low battery voltage

during CO

2

emergency cooling with

battery operation

Alarm message “

bt.E” immediately

LED indication immediately: LED

flashes, LED

flashes

Audible alarm: buzzer immediately

Switching the zero-voltage relay alarm output.

Actions:

Check the battery connections

Charge the battery (charging mode: current, approx. 160 mA)

If it is defective, replace the battery. Contact BINDER Service

Reset the alarm message with the “EXIT“ key.

Advertising