Ole_link19 – BINDER UF V 500 User Manual

Page 96

Advertising
background image

UF V (E2+E2.1) 03/2013

page 96/112

Battery alarm (no battery recognized)

chap. 10.6

Battery alarm (very low battery voltage)

chap. 10.6

Battery charging mode (information message, no alarm)

(low battery voltage, battery is charging)
This message can be displayed up to 45 hours with a new freezer or

after a power failure. With a freezer equipped with a CO

2

emergency

cooling (option) this message can be displayed up to 8 days.

chap. 10.6

Power failure alarm

chap. 10.7

Failure of Pt100 temperature sensors for the refrigerating machine

chap. 10.8

Failure of temperature sensor for ambient temperature

chap. 10.8

Failure of temperature sensor for safety controller

chap. 10.8

Failure of temperature sensor for interior measurement

chap. 10.8

Internal controller dysfunction

chap. 10.9

Internal controller dysfunction

chap. 10.9

Internal controller dysfunction

chap. 10.9

Internal controller dysfunction

chap. 10.9

CO

2

emergency cooling system (option) activated

chap. 11

Alarm CO

2

supply of emergency cooling system (option)

chap. 11

Alarm low battery voltage during emergency cooling with battery
operation

chap. 11

Early warning system cooling cycle

chap. 16

Ambient temperature alarm

chap. 16

Continuous operation of cooling cycle

chap. 16

Cooling system error

chap. 16

Condenser air filter and perhaps condenser soiled, or all ventilation slots

(at the front and bottom) are blocked.

chap. 14.4.1

Condenser fan defective

chap. 16

Advertising