11.3 Alarm messages

The alarm messages which occur simultaneously during operation with CO

2

emergency cooling are cycli-

cally displayed one after the other.
CO

2

emergency cooling system activated






Figure 57: CO

2

emergency cooling

system activated

Alarm message “

inJ” immediately

LED indication immediately: LED

flashes

Audible alarm: buzzer immediately

Switching the zero-voltage relay alarm output.

Note: Since the clocking of the CO

2

emergency cooling system always triggers the alarm again, you can-

not permanently reset the buzzer.

Insufficient CO

2

supply of the emergency cooling system






Figure 58: Alarm CO

2

supply of

emergency cooling system

Alarm message “

P.lo” immediately

LED indication immediately: LED

flashes

Audible alarm: buzzer immediately

Switching the zero-voltage relay alarm output.

Actions:

Check if the CO

2

gas supply is open

Connect a new CO

2

pressure cylinder

It is not possible to determine the filling level of a CO

2

pressure cylinder from the cylinder

pressure. Weigh the bottle regularly in order to replace it in time