10.8 Temperature sensor failure

A sensor fault alarm display takes priority over all other operational

displays and alarm signals on the controller.

Figure 49: Failure of Pt100 temperature sensor for the refrigerating machine. Refrigera-

tion is turned on permanently

Figure 50: Failure of Pt100 temperature sensors for the refrigerating machine

Figure 51: Failure of Pt100 temperature sensor for ambient temperature or air intake of

the 1

st

stage cooling (refrigerating machine)

Figure 52: Failure of Pt100 temperature sensor for safety controller: refrigeration is

turned on permanently

Figure 53: Failure of Pt100 temperature sensor for interior measurement: refrigeration is

turned on permanently

Alarm message “

992” or “997” immediately

LED indication immediately: LED

flashes

Audible alarm: buzzer (intermittent sound) immediately

Switching the zero-voltage relay alarm output.

Actions:

Please contact BINDER Service.


Alarm message “

992” or “998” or “999” immediately

LED indication immediately: LED

flashes

Audible alarm: buzzer (intermittent sound) immediately

Switching the zero-voltage relay alarm output.

The refrigerating machine is turned on permanently.

Actions:

Please contact BINDER Service.

If the temperature rises, i.e. the refrigerating machine is defective (safety controller temperature alarm

and / or tolerance range alarm):

Transfer the material to another freezer.

Turn off the freezer.

If necessary, clean and disinfect the freezer.