18.1 Factory calibration and adjustment

This unit was calibrated and adjusted in the factory. Calibration and adjustment were performed using

standardized test instructions, according to the QM DIN EN ISO 9001 system applied by BINDER (certi-

fied since December 1996 by TÜV CERT). All test equipment used is subject to the administration of

measurement and test equipment that is also a constituent part of the BINDER QM DIN EN ISO 9001

systems. They are controlled and calibrated to a DKD-Standard at regular intervals.

18.2 Over current protection

The devices are equipped with an internal fuse not accessible from outside. If this fuse is blown, please

inform an electronic engineer or BINDER service.

18.3 Battery replacement

Replace the battery only with a substitute of the same ratings (see chap. 18.6, art. no. 5007-0001). Re-

spect the correct polarity when replacing the battery!
Proceeding to exchange the battery:

Turn off the freezer at its key switch (6) and disconnect it from the power supply.

Remove the bottom cover on the rear of the unit

Remove the holding clamp of the battery

Disconnect the battery connections

When connecting the battery, respect the correct polarity.

When connecting the new battery, be sure to respect the correct polarity (red = positive, black = nega-

tive)

Secure the battery with its holding clamp

Reinstall the bottom cover on the rear of the unit

Connect the freezer to the power supply and turn it on with the key switch (6)