Overview of all controller indications, 1 indications of setting menus – BINDER UF V 500 User Manual

Page 94

Advertising
background image

UF V (E2+E2.1) 03/2013

page 94/112


Fault description

Possible cause

Required measures

Controller (continued)
Alarm test key cannot actuate the

alarm.

Main power switch (key switch) is

off.

Turn on the key switch (6).

Alarm test key actuates an in-

complete alarm (just message or

just buzzer).

When only the buzzer or only the

alarm is actuated by the alarm

test key, the part, which has not

been actuated is out of order and

must be replaced.

Contact BINDER service.

Automatic door mechanism

UF V (E2): The door cannot be

closed.

The door has been pressed on too

shortly (< 2 sec.) and was opened

immediately after. The automatic

door mechanism was triggered

and now prevents closing the door

again.

Contact BINDER service. Gener-

ally, press on the door for at least

2 seconds until the automatic

door mechanism is activated and

closes the door tightly.

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.


17. Overview of all controller indications

17.1 Indications of setting menus

Setting the temperature set point

chap. 6

Operating mode HAND

chap. 8.2

Operating mode USER

chap. 8.3

Locking and unlocking the operating functions’ settings, Operating mode

LOCK

chap. 8.4.1

Password request

chap. 8.3

Password for locking/unlocking

chap. 8.4

Password setting for access to operating mode USER

chap. 8.3

Setting the Safety controller set point (limit value)

chap. 9.1.1

Setting the address

chap. 8.3

Setting the ambient temperature alarm

chap. 8.3